To fix the global energy scale for the SRD measurement, I exploited the 3 calibration runs in which the 100 GeV muon beam impinged on each cell of this detector. Using this script, I evaluated the signal amplitude distribution and performed a fit with a Landau function convoluted with a Gaussian. The results are posted below. According to MC, the calibration constants are computed assuming that the mean energy deposition in the SRD is about 42 MeV.
RUN
SRD cell
MVP
10595
0
1404.1
10596
1
1155.4
10597
2
1093.4
Pietro Bisiochanged title from SRD calibration for 2024A to SRD energy calibration for 2024A
changed title from SRD calibration for 2024A to SRD energy calibration for 2024A
The red curve is the convolution between a Landau and a Gaussian function, used for the fit. The blue one is the Landau resulting from this fitting procedure.