@@ -8,7 +8,7 @@ From the RCMS interface, click on "Connect".
Open the [xDAQ](http://srv-s2g18-33-01.cms:20100/urn:xdaq-application:service=calibration) calibration suite. In case the calibration suite page was already opened before you clicked on "Connect" in RCMS, reload the page manually.
From the dropdown menu, select the calibration type "S-bit rate scan", then set the number of iterations in the "Iterations" text box to 1 and leave all the other parameters to their default values. Click on "Apply settings".
From the dropdown menu, select the calibration type "S-bit ARM DAC scan", then set the number of iterations in the "Iterations" text box to 6 and leave all the other parameters to their default values. Click on "Apply settings".
Go back to the RCMS interface, then follow these steps:
1. click on "Configure" and wait for all the subsystems to be in _Configured_ state;
@@ -17,6 +17,6 @@ Go back to the RCMS interface, then follow these steps:
Go back to the calibration suite in xDAQ, then click on "Run".
While the scan is ongoing, from the P5 DAQ machine check that the output files are being written to the subfolders `fed1467` and `fed1468` of the folder `/gemdata/runs/test/sBitRateScan/`.
While the scan is ongoing, from the P5 GEM DAQ machine check that the output files are being written to the subfolders `fed1467` and `fed1468` of the folder `/gemdata/runs/test/sBitRateScan/`.
## Analyzing the threshold scan results