GE11-P-36L1: HV sector shorted in G1Top, G3Top in etaP 2
Run: 381380
HV shorts in G1Top, G3Top.
For G1Top: when Vmon= 388V, Imon ~= 75uA
For G3Top, when Vmon= 364V, Imon ~= 72uA
This chamber is connected to GE11-P-34L2
Run: 381380
HV shorts in G1Top, G3Top.
For G1Top: when Vmon= 388V, Imon ~= 75uA
For G3Top, when Vmon= 364V, Imon ~= 72uA
This chamber is connected to GE11-P-34L2