Fast threshold scan might not start data taking or might run too long
The start of the data taking in the fast threshold scan can be overlooked when the stop condition is reached before data taking was started.
Also it can happen that not all enabled pixels reach maximum response even at maximal injection VCAL_HIGH_stop
(bad pixels or so). Then it pointlessly scans up to VCAL_HIGH_stop
although it could already stop when maximum response is reached.
Edited by Ghost User