Allow quick measurement of the DAC currents and voltages
Description
Recently, doubts about the DAC currents and voltages have recently been raised from time to time. Among others, during the "empty S-curves vs threshold investigations" and during the "empty S-curves and S-bit rate scan investigations". At the moment, one is forced to measure the DAC current and voltages by hand or to retake a complete DAC scan. This is long, error-prone, and/or useless.
Possible fixes
Develop a new tool, possibly a Python script, ideally a new feature in the AMCManger
. The feature is relatively tricky to implement in the current design of the software since calibration constants are exclusively used and available in the analysis suite. It might be easier after the cmsgemos
-cmsgemos-analysis
merge. Additionally, the AMCManager
version only makes sense when taking into account the most recent ideas about an Automator and the consequential Online Software restructuration.