Add GE1/1 VFAT FEAST voltage monitoring
Description
The title should speak for itself.
Also includes various bug fixes or enhancements related to the SCA-based monitoring.
- Ensure that the ADC conversions are over before retrieving the measurement, even in the case of an AXI bus error.
- Make use of the 1kOhm precision resistor to calibrate the ADC current source (improve dramatically the PT100/PT1000 precision).
- Export the raw ADC counts in the JSON monitorable object for debugging purposes.
Related Issue
None...
How Has This Been Tested?
Thoroughly tested on the GE1/1 QC7 setup.
Types of changes
-
Bug fix (non-breaking change which fixes an issue) -
New feature (non-breaking change which adds functionality) -
Breaking change (fix or feature that would cause existing functionality to change)
Checklist:
-
My code follows the code style of this project. -
My change requires a change to the documentation. -
I have updated the documentation accordingly. -
I have read the CONTRIBUTING document. -
I have added tests to cover my changes. -
All new and existing tests passed.
Edited by Laurent Petre