New instructions for GEM FED 1467
There are new special instructions for GEM FED 1467. They have been added to the shift bulletin board. It would be nice if the DAQExpert could also display them:
- If GEM are OUT_OF_SYNC -> Perform the hard reset without calling the GEM DOC.
- If GEMs give events out of sequence (RU is in sync loss) -> stop/start the run, then call the GEM DOC.
- If GEMs require green recycle -> call GEM DOC BEFORE performing it. GEM DOC will take debug information, then proceed with the green recycle.
- During collisions, if GEMs stop the run more than once, they are to be immediately taken out (notify the GEM DOC after).
- If the GEM FM goes into ERROR with the message "...OptoHybrid connected on link X to AMC in slot Y is not responding", please call the GEM DOC, as this indicates a communication failure with the front-end and likely requires an LV power-cycle to be performed by the GEM DOC.
The last point is old and might already be implemented in the DAQExpert.