ItkPixV1.1: threshold distributions shifted by -100/200e from the target threshold
as discussed on mattermost:
somewhat a follow-up of #68 (closed) for ITkPixV1.1, where the situation looks worse.
we see systematically a lower threshold than the target by 100 or 200+ electrons when the target is 1000e. It may be an issue with the threshold tuning algorithm?
A few example plots are below. All these chips are in the PDB, chip config files can be accessed from here: https://gitlab.cern.ch/SaclayITk/yarr-apps/module-scans (just request access)